48-7009-E 6-6
PV X-Y 2-POINT CORRECTION SETTING
Determine any two points within the input range (upper and lower limit), then convert to corrected value to the PUV of Input 1.
For Example)
When the PV of the product is;
90 indication but actual temp. is 100
310 indication but the actual temp. is 300
If these are to be corrected by 2-point Correction Setting;
PV before correction: PX 1 = 90, PX 2 = 310℃
PV after correction: PY 1 = 100, PX 2 = 300℃
PX 1 PV Before Correction Lower Limit Value Setting
Setting RangeSetting Range Lower Limit to (PX 2 1) or Setting Range Lower Limit to (PX 2 1.0)
Setting Range Lower Limit to (PX 2 10-digits)
Setting Units1 or 0.1, 1-digit
PX 2 PV Before Correction Upper Limit Value Setting
Setting Range(PX 1 + 1) to Setting Range Upper Limit , or (PX 1 + 1.0) to Setting Range Upper Limit
(PX 1 + 10-digits) to Setting Range Upper Limit
Setting Units1 or 0.1, 1-digit
PY 1 PV After Correction Lower Limit Value Setting
Setting RangeSetting Range Lower Limit to (PY 2 1) or Setting Range Lower Limit to (PY 2 1.0)
Setting Range Lower Limit to (PY 2 10-digits)
Setting Units1 or 0.1, 1-digit
PY 2 PV After Correction Upper Limit Value Setting
Setting Range(PY 1 + 1) to Setting Range Upper Limit , or (PY 1 + 1.0) to Setting Range Upper Limit
(PY 1 + 10-digits) to Setting Range Upper Limit
Setting Units1 or 0.1, 1-digit
NOTE: - As much as possible, set the difference in the value of Before Correction and After
Correction bigger within the realizable range. When the difference is small, the margin of
error might be bigger.
- When the slope of PX 1 and PX 2 (before correction) is set to 1 and when the slope after
Correction becomes less than 1, the indication range in each input type becomes narrow.
PX1
PY1
PY2
PX2
After Correction
Before Correction
PV値
Actial Input
Upper Limit Value
Lower Limit Value